Current¶
SCPI Commands
READ:EVDO:MEASurement<Instance>:MEValuation:TRACe:EVMagnitude:CURRent
FETCh:EVDO:MEASurement<Instance>:MEValuation:TRACe:EVMagnitude:CURRent
- class Current[source]
Current commands group definition. 2 total commands, 0 Sub-groups, 2 group commands
- fetch() → List[float][source]
# SCPI: FETCh:EVDO:MEASurement<instance>:MEValuation:TRACe:EVMagnitude:CURRent value: List[float] = driver.multiEval.trace.evMagnitude.current.fetch()
Returns the values of the RMS EVM traces. A trace covers a time interval of 833 μs (one half-slot) and contains one value per chip. The results of the current, average and maximum traces can be retrieved.
Use RsCmwEvdoMeas.reliability.last_value to read the updated reliability indicator.
- return
curr_evm: float Range: 0 % to 100 %, Unit: %
- read() → List[float][source]
# SCPI: READ:EVDO:MEASurement<instance>:MEValuation:TRACe:EVMagnitude:CURRent value: List[float] = driver.multiEval.trace.evMagnitude.current.read()
Returns the values of the RMS EVM traces. A trace covers a time interval of 833 μs (one half-slot) and contains one value per chip. The results of the current, average and maximum traces can be retrieved.
Use RsCmwEvdoMeas.reliability.last_value to read the updated reliability indicator.
- return
curr_evm: float Range: 0 % to 100 %, Unit: %